Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
139
7.5.6
Setting limits
Two limits for monitoring the measured value can be set for each measured component and
auxiliary value. The limit message can be triggered either with high or low violation of the
limit. It is also possible to set the switching hysteresis and the category to which the
message is to belong ("
Signal type"
).
{"Parameter";"Limit setting"}
Menu call:
Main menu → Parameters → Limits → [Measured component] → Limit [n]
The following menu is displayed (example):
You can then set the individual parameters as follows:
Setting a limit
Use the
,
and
keys to select the first parameter "Value".
Set the desired limit in the follow-up menu as described in section Character input (Page 64).
Selecting the message type
Use the
,
and
keys to select the second parameter "Type".
Set the message type in the follow-up menu using the
,
and
keys. You can
select from the following:
●
(+) High violation
and
●
(-) Low violation.
Hysteresis
Use the
,
and
keys to select the third parameter "Hysteresis".
Set the desired value in the follow-up menu as described in section Character input
(Page 64).