Operation
6.1 Control panel, display and control elements
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
57
6.1.4
Messages in the status line
Various status messages can be displayed in the status line of the SIPROCESS UV600, as
listed in the following table:
Table 6- 2
Messages in the status line
Message
Description
Possible measures
Warm-up
The analyzer has not yet reached
its operating temperature
⇒ Wait until "Measuring" is displayed.
☹
The measured values of the analyzer could not
yet be used.
☹
Do not carry out calibrations
☺
Other operating functions can be carried out.
Measuring
The analyzer module is in the
normal operating state
Only appears if the menu branch of an analyzer module is
called.
Purging
A purging time is active, measuring
mode is interrupted
Cal/Val […]
A calibration procedure or
validation measurement is
currently running
[current phase]
⇒
If a calibration procedure or manual validation is being carried
out:
Continue the procedure in accordance with the instructions (see
Auto-Hotspot).
⇒ If a fully automatic calibration or validation is being carried out:
Wait until the automatic procedure is finished.
See logbook
A status or function message has
occurred
⇒ If the reason is unknown: Check the logbook (see
)
.