Important functions (selection)
7.2 Diagnostics functions
SIPROCESS UV600
82
Operating Instructions, 10/2013, A5E03529375-02
The following detailed information can be viewed:
Notation in the menu
Explanation
NO
Assigned measured component
Current
Actual output value
Zero
Live zero (zero of output current range)
Range sel.
Mode of output range selection
Status (AOnR)
Current output range (0/1) [AOn = assigned analog output]
Range 0
Activation status of output range 0 (on/off)
Range 0 Start
Physical start-of-scale value of output range 0
Range 0 End
Physical full-scale value of output range 0
Range 1
Output range 1: same parameters as range 0
Range 1 Start
Range 1 End
Digital outputs
Menu call:
Main menu → Diagnostics → IO → Digital output
The following menu is then displayed (example):
The menu displays all digital outputs which have a function, together with their current
statuses (0 = inactive, 1 = active). You must take into account the set switching logic (see
menu 5.7.2 in section Configuring the digital outputs (Page 145)). In order to obtain detailed
information on a digital output, select it using the
and
keys and then by pressing
the
key.