Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
148
Operating Instructions, 10/2013, A5E03529375-02
Assigning the internal function (destination) to a digital input
In the previous menu, select the digital input. In the subsequent menu, select 'Destination'.
The following menu is displayed (example):
The parameters have the following meanings:
Name of calibration
Programmed automatic calibration
Name of validation
Programmed automatic validation measurement
Cancel Cal./Val.
Cancel a running calibration/validation
Failure
Activate status flag "F" (for complete system)
Maintenance
Activate status flag "M" (for complete system)
Pump off
External control signal for switching off the gas pump
Calibration gas fault
External fault message "Calibration gas not available"
[1]
Disable Cal./Val.
External control signal to prevent calibrations/validations
[1]
None
Not occupied (electronic status of digital input has no effect)
[1]
Prevents calibrations and validations
1.
Select the desired destination using the
,
and
keys.
2.
Search for the desired function which the digital input is to control in the follow-up menus.