Mounting and connecting
4.5 Electrical connections
SIPROCESS UV600
30
Operating Instructions, 10/2013, A5E03529375-02
4.5.1
Power connection
4.5.1.1
Safety instructions for power connection
WARNING
Missing PE/ground terminal
Risk of electric shock!
Only connect the device to a power supply which has a functional ground conductor (PE,
protective earth).
The PE conductor must not be interrupted inside or outside the device.
Depending on the device version, connect the power supply as follows:
•
Power plug:
Ensure that the socket used has a PE/ground conductor connection. Check that the
PE/ground conductor connection of the socket and power plug match each other.
•
Connecting terminals:
Connect the terminals according to the terminal assignment table. Connect the
PE/ground conductor first.
NOTICE
Faulty power supply
Device failure with faulty power supply
•
Check that the existing power supply agrees with that specified on the rating plate. If this
is not the case, the device must not be connected or started up.