Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
111
The calibration procedure commences. The individual phases are displayed on the screen
during execution of the calibration procedure. The meanings are as follows:
Phase
Text in the status line
1 - Flow of calibration gas
Purging with cal. gas
2 - Determination of measured value of calibration gas (actual value)
Measuring
3 - Validation: the results are calculated and saved
Calibration: calibration running
Calculating
4 - Flow of sample gas
Purging with sample gas
The following menu is displayed (example):
The phase is displayed in the status line, the other parameters are as follows:
Display on screen
Description
1 Remaining time
Remaining time for current phase
2 NO
Component with current measured value (actual value)
3 Mean value
Mean value of measured values (during the 'Measuring' phase)
4 Zero gas
Name of calibration gas with setpoint
5 Cancel (not shown here) Cancel current procedure