132
© Servomex Group Limited. 2020
07931001B revision 6
Safety
Do not install the Laser 3 Plus in a high velocity dust-laden atmosphere.
You are responsible for ensuring that:
•
The instrument is installed correctly and safely
•
The laser beam is enclosed during operation
•
The sampling system is leak free (if being used in bypass
extractive configuration)
•
The venting system is appropriate for the gases you are sampling
•
The installation does not introduce trip hazards
•
Appropriate PPE is used for installation, servicing and
decommissioning
Lubricate flange mounting threads with supplied anti-seize paste before
you install the Laser 3 Plus.
The optical windows in the receiver and transmitter units are installed in
the factory. Do not touch the optical windows as any contamination will
adversely affect the performance of the Laser 3 Plus. For cleaning
guidance see section 8.3.
Regularly inspect, test and replace seals to ensure that all external
connections are always leak-free at full operating pressure.
The equipment is incapable of passing the dielectric strength test
prescribed by the standards, and so this must be taken into account
during installation by using an SELV power source with a prospective
short circuit current not exceeding 40A.