88
© Servomex Group Limited. 2020
07931001B revision 6
Note:
“Target” refers to the target concentration of the calibration gas
Note:
“Tolerance” refers to the tolerance of the measured reading against
the
“Target” concentration. A warning will be indicated if the measured
concentration value is outside the “Tolerance” when asked to confirm
calibration.
Hint:
You may wish to check the raw signal and filtered graphs to check
correct location of second harmonic signal and 2F burst
Basic cal measurement type: Uses raw transducer measurement, prior to any X-
interference compensation or measurement filtering.
X- interference cal measurement type: Uses the X-interference compensated
measurement prior to any measurement filtering.
Note
: X- interference must be enabled in set-up.
Note
: X-
interference requires “Supervisor” password or above
.
Settings
Cal Limits
1 Xxxx Limits
Cal Values
1 XXXX Measurement
Min Cal Target Diff
Cal Limits
2 XXXX Measurement
Cal Measurement Type
0.00%
edit value
Settings
Cal Measurement Type
1. Xxxx Cal Meas Type
Cal Values
1 XXXX Measurement
Cal Measurement Type
Cal Limits
2 XXXX Measurement
Filtered Measurement
Cal Measurement Type
Basic Measurement
select
X-Interference