07931001B revision 6
© Servomex Group Limited. 2020
69
4.11 Gain and phase settings
DC Gains 1 and 2 enable the signal to be optimized for the specific application; in
addition, the phase of the measurement signal and reference burst must also be
optimized. The gain setting may need to be changed due to long pathlengths, dust,
and beam divergence which can reduce the amount of laser energy reaching the
receiver.
To view photodiode DC Level
The DC photodiode level tells the user how much laser energy is being received by
the detector. Setting it to a specific level (by adjusting alignment and gain) can help
improve measurement accuracy and ensure transmission is maintained to an
acceptable level.
Figure 4-31: Gain and phase setting menu
DC Gain 1 and DC Gain 2 are typically set for each application and may need to be
changed during commissioning to obtain optimum performance.
Note:
If the photodiode DC level is too high an error status will be raised and
the analyzer will report incorrect readings.
The photodiode DC level (DetSigV) must not exceed 3.5V DC
.
In dusty processes, where the dust loads vary significantly, gain levels
set when dust loads are either very high, or very low can cause zero
detection (not enough laser energy) or saturation (too much laser
energy) when dust levels revert to normal. Review the application and
dust load variations and consult Servomex for advice on how to
maximize operation of the analyzer in these situations.
Note:
Analyzer alignment is critical to obtaining a reliable measurement.
Excessive gain levels used to compensate poor alignment can cause
the measurement to become excessively noisy and reduce the LDL.
See sections 7.3.1and 7.3.2.
Note
You must use ‘Adjust Burst Phase Cal’ after changing the DC Gain
. Do
so with caution where process temperatures are above 900°C or gas
concentrations in the process are high. See section 4.11.4.
Level 1
Level 2
Level 3
Level 4
Level 5
Menu
Set Up
Manage Analyser
Manage Current Config Phase/Gain Settings
Calibrate
Manage Analyser
Manage Current Config
Service Mode
Alignment LEDs
Data Log
mA Input
Save Current Config
Phase/Gain Settings
Alignment Lock Level
Alarms
mA Output
Restore Config
Measurement Signal
Photodiode DC level
Status
Relay Setup
Delete Config
Physical Setup
DC Gain 1
Set Up
Filtering
Detailed Setup
DC Gain 2
Measurements Unit Select
Transmission Level
Service
X-Interference
Raw Signal
Clipping
AC Amplifier Gain
User Settings
ABurst
Adjust Burst Phase Cal
Ref TransCal
Summary of Contents for 07931001B
Page 2: ...Servomex Group Limited 2020 This page intentionally blank...
Page 4: ...Servomex Group Limited 2020 This page intentionally blank...
Page 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Page 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...