All Good Path
Failure Path
9
Enter the tolerance of the validation
gas. This will be the tolerance on the
validation gas bottle.
NOTE THIS FEATURE IS
DEPRECATED AND THIS VALUE
SHOULD BE MAINTAINED AS
EQUAL TO ZERO.
Uncertainties related to the gas
bottle can alternatively be set at step
number 10
10
Enter the permitted tolerance on the
span concentration calculated by the
analyzer.
This is expressed as a percentage of
the span concentration. For
example, a Validation Pass Level of
5%, with a 21%O2 validation test
gas bottle would correspond to
successful validation if the reported
validation readings would be
between 20%O2 and 22%O2.
11
Enter the length of the validation
cell.
Standard lengths for Servomex
supplied in-line validation cells are
either 0.1m or 0.2m.
You cannot enter a number <0.1m
12
Enter the permitted stability of your
process concentration expressed as
percentage standard deviation of the
span concentration.
At all stages of the validation
process, the analyzer will wait for the
measurement concentration to
achieve this stability before
proceeding refer to section 6.12.3 for
more details of the process stability
measurement.
13
Enter the permitted stability of your
process transmission expressed as
percentage standard deviation.
Summary of Contents for 07931001B
Page 2: ...Servomex Group Limited 2020 This page intentionally blank...
Page 4: ...Servomex Group Limited 2020 This page intentionally blank...
Page 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Page 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...