6. Calibration
Item
Ref
Checked
1. Save the configuration settings
2. Remove the transmitter and receiver units from the
process
3. Connect transmitter and receiver units to the calibration
cell
4. Reconfigure settings for offline calibration
5. Save the calibration configuration settings
6. Calibration settings
7. Calibrate and verify correct concentration is displayed
8. Saving calibration configuration
9. Restore configuration settings
Table 6-1: Calibration check list
The analyzer is factory calibrated using a certified gas mixture and is supplied with a
calibration test report.
Servomex recommend that you verify the calibration of the instrument annually,
using a certified test gas and the supplied calibration gas cell.
6.1 Save the configuration settings
Save all configuration data as described in sections below:
Measurement calibration configuration
Analyzer configuration (User settings) Saving/ loading
Note:
Save and store in a safe place with date and description so that the files
can be used for future reference.