07931001B revision 6
© Servomex Group Limited. 2020
55
Note:
The actual mA output value is controlled from the mA output calibrate screen as
long as the screen is displayed. As soon as the mA service screen is no longer
displayed, the mA output value will be updated to reflect the corresponding
measurement unless the corresponding measurement as stated is jammed high
or low.
4.5 Configure measurement alarms
To view active measurement alarm status:
Figure 4-15: View active measurement alarm status menu
Alarm modes and levels
25 alarms are available to be allocated for the gas measurements or transmission.
You can configure each alarm to operate in one of three modes:
Alarm mode Operation
None
An alarm is raised when a sample measurement is
lower than the pre-set alarm level
Low alarm
An alarm is raised when a sample measurement is
lower than the configured alarm level
Note:
During calibration an alarm is only activated if
you have set the alarm ‘Follow’ option to Yes
High alarm
An alarm is raised when a sample measurement is
higher than the pre-set alarm level
Note:
An alarm is only activated during calibration if
you have
set the alarm ‘Follow’ option to Yes
Alarms
Active Alarm
Active
Set Up
Follow
None
History