07931001B revision 6
© Servomex Group Limited. 2020
19
2.8 Materials in contact with the sample
Item:
Material:
Flange:
316 stainless steel
Insertion tube:
316 stainless steel
O-rings:
Fluorocarbon
Chemraz 505
(optional)
Process flange gasket:
Stainless steel and graphite composite
Optical windows:
Fused silica (optical coating)
Sapphire (optical coating) (
optional)
Optical window seal:
Loctite 595
2.9 Unpacking
Remove the SERVOTOUGH Laser 3 Plus and its accessories from the packaging
and inspect everything for any damage that may have occurred during transit.
If any item has been damaged, contact Servomex or its agent straight away.
Keep all packaging and shipping information.
Check that the parts supplied agree with your purchase specification.
Wetted materials must be compatible or clean for oxygen service at
concentrations above 21% or equivalent at elevated pressure.
Remove Silica Gel sachet before attempting to operate the
instrument.