4.3 Settings Related to Process Conditions
“
Physical Set-
up”
To provide an accurate measurement, process pressure, temperature and the
length where the measured gas will occur should be known and correctly set into the
analyzer.
Configuration includes setting the laser path length, process pressure and process
temperature etc.
Note:
When the laser measurement signals have been optimized and the
process environment set up has been completed Servomex recommend
saving the configuration. See section 0
The physical parameters that must be set up are illustrated in the following
sequential menu table:
Figure 4-10: Physical setup sequential menu
Menu
Manage Analyser
Calibrate
Level 3
Level 4
Level 5
Level 6
Level 7
Level 8
Level 9
Data log
Alarms
Status
Set up
Measurement
Service
Enabled
Disabled
edit value
Delete Config
edit value
Detailed set up
For Servomex trained Service operators only
Save Current Config
Restore Config
Proc:Temp. Offset
Offset or value
Edit
select option
select option
Proc:Pressure Source
Proc:Temperature Source
mA Input 1 or 2
select option
Receiver Enclosure
Proc:Pressure Source
Offset or value
Edit
Transmitter Enclosure
edit value
Physical Setup
Measurement
Selection
Purge Compensation
Proc:Path Length
0.000
Transmitter Enclosure
User defined 0-4
mA Input 1 or 2
User defined 0-4
Manage Current Config
Service Mode
Phase/Gain Settings
Measurement Signal
Summary of Contents for 07931001B
Page 2: ...Servomex Group Limited 2020 This page intentionally blank...
Page 4: ...Servomex Group Limited 2020 This page intentionally blank...
Page 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Page 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...