![Servomex 07931001B Operator'S Manual Download Page 163](http://html1.mh-extra.com/html/servomex/07931001b/07931001b_operators-manual_1247057163.webp)
07931001B revision 6
© Servomex Group Limited. 2020
149
Ethernet connector
Figure 7-20: Ethernet connections
RJ-45 Pin
T-568A Wire Color
T-568B Wire Color
Signal 07900922 Connector Pin
1
Green/White
Orange/White
Tx +
2
2
Green
Orange
Tx -
1
3
Orange/White
Green/White
Rx +
4
4
Blue
Blue
-
5
Blue/White
Blue/White
-
6
Orange
Green
Rx -
3
7
Brown/White
Brown/White
-
8
Brown
Brown
-
A 4-way Weidmuller plug is used hence a modified Ethernet cable is required to
connect to the instrument. The table below shows the two standards of wiring RJ-45
connectors and the corresponding signal names and 07900922 connections
Note:
The instrument supports Auto-MDIX (auto cross-over) so you can use either a
straight or cross-over cable.
Hint:
Installation of an Ethernet cable to a designated area suitable for remote
connection to a PC is recommended.