- 21 -
5.
Linkage between remote alarm and buzzer
Setting of linkage between remote alarm and buzzer
“0” --- Remote alarm is inactive by pressing BUZZER key.
“1” --- Remote alarm isn’t inactive by pressing BUZZER key.
Contact output
Between N.O. and COM.
Between N.C. and COM.
At normal
Open
Close
At alarming or power failure
Close
Open
6. Error
messages
Condition Display
High temperature alarm
“Warning:High Temp”
Low temperature alarm
“Warning:Low Temp”
High CO2 density alarm
“Warning:High CO2 Density”
Low CO2 density alarm
“Warning:Low CO2 Density”
High O2 density alarm
“Warning:High O2 Density”
Notice of UV lamp life
“Warning:UV Life”
CO2 gas becomes empty
“Err01:CO2 Gas Empty”
N2/O2 gas becomes empty
“Err02:N2/O2 Gas Empty”
Wrong connection of CO2 line
“Err03:CO2 Gas Connect Abnormal”
Wrong connection of N2/O2 line
“Err04:N2/O2 Gas Connect Abnormal”
Temp. sensor is disconnected
“Err05:Temp Sensor Open”
Temp. sensor is short circuited
“Err06:Temp Sensor Short”
CO2 sensor box temp sensor is disconnected
“Err07:CO2S Box Temp Sensor Open”
CO2 sensor box temp. sensor is short circuited
“Err08:CO2S Box Temp Sensor Short”
Ambient temp. sensor on PCB is disconnected
“Err09:AT Sensor Open”
Ambient temp. sensor on PCB is short circuited
“Err10:AT Sensor Short”
CO2 sensor Vref voltage is abnormal
“Err11:CO2 Sensor Vref Abnormal”
CO2 sensor Vgas voltage is abnormal
“Err12:CO2 Sensor Vgas Abnormal”
Main heater is disconnected, SSR is short circuited
“Err13:Main Heater Abnormal”
Bottom heater is disconnected, SSR is short circuited “Err14:Humidity Heater Abnormal”
Door heater is disconnected, SSR is short circuited
“Err15:Door Heater Abnormal”
CO2 sensor box heater is disconnected, SSR is short
circuited
“Err16:CO2S Box Heater Abnormal”
SSR for each heater is disconnected
“Err17:Heater SSR Open”
UV lamp malfunction
“Err18:UV Lamp Abnormal”
O2 sensor malfunction
“Err19:O2 Sensor Abnormal”
H2O2 decontamination kit is unconnected
“Err31:H2O2 Unit No Connect”
H2O2 level is low, water sensor malfunction
“Err32:H2O2 Level”
Door open at starting at H2O2 decontamination
“Err33:Door is Opening”
H2O2 volume is abnormal
“Err34:H2O2 Volume”
Power is interrupted during H2O2 decontamination
“Err35:Power Failed”
UV lamp malfunctions during UV resolution
“Err36:UV Lamp Failure”
Power is interrupted during UV resolution
“Err37:Power Failed”
Standard gas becomes empty
“Err41:Std Gas Empty”
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...