ALARM PARAMETERS
1.
Press the MENU Key from the Top Display to display the menu, select
Tools,
and press the ENTER
Key.
2.
Select
Alarm Setting
from the Select Tools Display, press the MENU Key to display the menu, select
OK
, and press the ENTER Key.
3.
The Alarm Setting Display will appear. The temperature alarm, CO
2
alarm, O
2
alarm, alarm delay, ring
back time, and door delay can be set on this display. The alarm buzzer can be silenced by pressing the
BUZZER Key. The buzzer will sound again after the specified ring back time if the condition that caused
the alarm continues. The ring back time can be set.
・
Temperature alarm: ±1.0 to ±5.0
o
C (Default: ±1.0
o
C)
・
CO
2
alarm, O
2
alarm: ±0.5 to ±5.0% (Default: ±1.0%)
・
Alarm delay: 0 to 15 minutes (Default: 15 minutes)
・
Ring back time: 1 to 99 minutes, or OFF (Default: 30 minutes)
・
Door alarm delay: 1 to 30 minutes (Default: 2 minutes)
0
1
0
2
0
3
0
4
0
5
0
6
0
7
0
8
0
9
1
0
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
2
0
2
1
2
2
2
3
2
4
2
5
2
6
2
7
2
8
M E NU
O K
C a n ce l
2
3
0
3
1
3
2
3
3
3
4
3
5
3
6
3
7
3
8
3
9
4
0
S
e l e c
t T
o
o
l s ( 1 / 2 )
T e m p / C O 2 / O 2 C a l i b r a t i o n
A l a r m
S e t t i n g
L
C
D
/ D
A
Q
S
e t t i n g
U V S e t t i n g
K e y L o c k P W S e t t i n g
D a t e T i m e
M E N U
O K
S
v c
C
a n
c e l
0
1
0
2
0
3
0
4
0
5
0
6
0
7
0
8
0
9
1
0
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
2
0
2
1
2
2
2
3
2
4
2
5
2
6
2
7
2
8
2
9
3
0
3
1
3
2
3
3
3
4
3
5
3
6
3
7
3
8
3
9
4
0
1
A l a r m S e t t i n g
3
T
e m
p
A
l a r m
±
1 . 0
o
C ( ± 1 . 0 ° C
-
± 5 . 0 ° C
)
C
O 2
A
l a r m
± 1 . 0 % ( ± 0 . 5 %
-
± 5 . 0 %
)
4
O 2
A l a r m
±
1 . 0 %
( ± 0 . 5 %
-
± 5 . 0 %
)
5
A
l a r m D e l a y 1 5 m i n ( 0 - 1 5 m i n )
R i n g B a c k 3 0 m i n ( 0 . O F F 1
-
9 9 m i n
)
6
D o o r D e l a y
2 m i n
( 1 - 3 0 m i n )
M E N U
O K
C a n c e l
5
.
1
T e m p
3 7 . 0
o
C C
O
2
5 . 0 %
O 2 5 . 0
%
2
3
4
5
S
t a t u s
6
O K
7
37
.
0
5
.
0
5
.
0
M E N U
S
e t
L o g
T o o l s
- 140 -
33
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...