1.
Outer door:
The outer door is held to the frame with a magnetic seal. A door heater is installed in the
door panel. The door opening is reversible. Contact a Sanyo representative or agent to change the door
hinge from left to right or vice versa.
2. Inner door:
Open when take the trays out or clean inside.
3. Gastight split door:
Made of tempered glass, however avoid excessive impact on the glass.
4. Leveling feet:
The leveling feet can be turned to adjust the height. Adjust the feet so that the Incubator
is level.
5. Trays:
Set trays at the same level as each gastight split door. Open a inner door when take the
tray out.
6. Tray supports:
The tray supports can be removed by lifting the front side and pulling toward you.
7. Side supports:
The right and left side supports can be removed for disinfection. Refer to page 45.
8. Duct:
The duct for the path for circulating air. It is removable.
9. Fan cover:
The fan cover serves as the inlet for circulating air. It is removable.
10. Fan (inside the duct):
The fan is made from polypropylene resin. It can be disinfected in an
autoclave.
11. Sample air outlet:
The sample air outlet also functions as an internal gas outlet. Normally, cover this
outlet with the sample air outlet cap.
12. Sample air outlet cap:
Always attach this cap except when using the sample air outlet.
13. Door switch:
Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
2
and N
2
/O
2
when door is open. UV lamp is also deactivated by door opening*.
14. Humidifying pan:
Fill the humidifying pan with sterile distilled water. Install the pan properly so that it
can be covered with the pan cover.
15. Humidifying pan cover:
The cover prevents UV light entering the chamber. When filling the pan, lift
the front side and take out the pan. Refer to page 48 for details.
16. UV lamp*:
This Sanyo UV lamp does not generate ozone. Never look directly at the UV light. For
replacement, contact a Sanyo representative or agent.
17. Water level sensor for humidifying pan:
This sensor detects the water level in the humidifying pan.
Refer to page 44 for details.
18. Remote alarm terminals:
Refer to page 13.
19. Access port:
Place the silicon caps on the port both outside and inside when the port is not being
used.
20. Power switch:
This is the main switch for the Incubator. It also functions as an overcurrent breaker.
21. Glow starter*:
The glow started if for the UV lamp (model FG-7P)
22. Connecting port A/B for CO
2
gas pipe:
When the optional MCO-21GC Automatic CO
2
Cylinder
Changeover System is installed, both ports A and B are available. If the MCO-21GC is not installed, only
port A is available. Refer to page 16 for gas cylinder connection. Ensure that the gas pressure is set at
0.03 MPaG (0.3 kgf/cm
2
G, 4.3 psiG). Refer to page 67 for automatic CO
2
cylinder changeover.
- 117 -
10
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...