1. In Top Screen, press MENU key and select “Tools”.
2. Press ENTER key to display “Select Tools (1/2)” and select “Temp/CO2/O2 Calibration”.
3. Press ENTER key to display “Temp/CO2/O2 Calibration (1/2)”.
4. Input a O2 measured value in the right side of “O2 Span”.
5. Press MENU key and select “OK”.
6. Press ENTER key to start O2 calibration automatically.
㻌 㻌 㻌
Perform O2 calibration until display CO2 density corresponds to actual CO2 density.
Note) It is impossible to perform O2 span calibration when setting O2 density deviates from 18.0~22.0%.
There is a blank in the right side of “O2 Span” column.
Ex.
䠅
When O2 density setting is 5.0% and actual O2 density is 4.5%;
Procedures MENU
key
0
Measure O2 density by sampling port in the front of inner door.
1
In Top Screen, press MENU key and select “Tools”.
Tools
2
Press ENTER key to display “Select Tools (1/2)” and select “Temp/CO2/O2 Calibration”.
3
Press MENU key and select “OK”.
OK
4
Press ENTER key to display “Temp/CO2/O2 (1/2)”. Input “4.5” in the right side of “O2
Span” column.
5
Press MENU key and select “OK”.
OK
6
Press ENTER key to return the display to “Select Tools (1/2)”.
㻌
㻌
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...