- 46 -
㻌
㻌
㻌
㻌
<Setting of CO2 density for standard gas calibration>
Setting of CO2 density for standard gas calibration (optional setting)
Setting range is 0.0%, 4.0~21.0%. (Factory default: 5.0%)
When you set “0.0”, CO2 using by standard gas will not be calibrated.
Note) When you input a value lower than “4.0”, CO2 density for standard gas calibration is automatically set in 4.0%.
When you input a value higher than “21.0”, CO2 density for standard gas calibration is automatically set in 21.0%.
<Setting of O2 density for standard gas calibration>
Setting of O2 density for standard gas calibration (optional setting)
Setting range is 0.5~81.0%. (Factory default: 5.0%)
When you set a value between 18.0~22.0%, O2 using by standard gas will not be calibrated.
Note) When you input a value lower than “0.5”, O2 density for standard gas calibration is automatically set in 0.5%.
When you input a value higher than “81.0”, O2 density for standard gas calibration is automatically set in 81.0%.
Ex) When both CO2 and O2 are set in 5% in standard gas calibration;
Procedures MENU
key
0
In Top Screen, press MENU key and select “Tools”.
Tools
1
Press ENTER key to display “Select Tools (1/2)”.
Press down shift key (
䊼
) to display “Select Tools (2/2)”.
2
Select “Standard Gas Setting” in “Select Tools (2/2)” display.
3
Press MENU key and select “OK”.
OK
4
Press ENTER key and input “5.0” in the right side of “Std Gas Co2 Density” column in
“Standard Gas Setting” display.
5
Press ENTER key and input “5.0” in the right side of “Std Gas O2 Density” column in
“Standard Gas Setting” display.
6
Press MENU key and select “OK”.
OK
7
Press ENTER key to return to “Select Tools (1/2)” display.
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...