6.
When gas cylinder A is switched to B, remove gas cylinder A and replace it with new one.
Note:
Be careful when handling the empty CO
2
gas cylinder. Some gas may still be left in the cylinder.
Ɣ
When gas cylinder B becomes empty, the line is switched to gas cylinder A again.
The MCO-21GC detects that there is no more CO
2
gas in a cylinder when the CO
2
density in the chamber
does not increase for a while even after opening the CO
2
gas valve in the Incubator. The supply line may
also be switched for other reasons even though there is gas left in the cylinder: blocking or restricting of
gas tube, reduction of CO
2
gas pressure, or improper opening of CO
2
gas cylinder. Always check the gas
quantity in the cylinder before disconnecting it.
Ɣ
Use the following procedure to manual switch the line between CO
2
cylinders A and B (only when
MCO-21GC is installed). This procedure shows switching the line from
CO
2
cylinder B to A as an
example.
1.
From the Top Display, press the Left Cursor Key for 3 seconds.
2.
The CO
2
gas supply line will be switched from cylinder B to A
(When the line needs to be switched from cylinder A to B, press the Right Cursor Key for 3 seconds.)
5
.
1
T e m p
3 7 . 0
o
C C O 2
5 . 0 % A
B
O 2 5 . 0
%
A
B
2
3
4
5
S
t a t u s
6
O K
7
37
.
0
5
.
0
5
.
0
5
.
1
T e m p
3 7 . 0
o
C C O 2
5 . 0 %
A
B O 2 5 . 0
%
A
B
2
3
4
5
S
t a t u s
6
O K
7
37
.
0
5
.
0
5
.
0
- 175 -
68
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...