Temperature calibration
㻌
Here is the explanation how to calibrate chamber temperature.
Ex) When setting temperature is 37.0
䉝
and actual temperature is 36.5
䉝
Process Operation MENU
0
Set thermal sensor at the center of the chamber and start the unit running.
Leave it until chamber temperature is stable.
1
Press MENU key in Top screen to call MENU screen. Select Tools.
Tools
2
Select Tools (1/2) screen is displayed. Select Temp/CO
2
Calibration.
3 Temp/CO
2
Calibration is displayed and press MENU key to activate MENU key. Select
OK and press ENTER key.
OK
4
Change the value in the column Temp Span to
36.5
䉝
by pressing digit shift key and
numerical value shift key.
5
Press MENU key to call MENU screen. Select OK and press ENTER key.
OK
6
Automatically reverts to Select Tools
㻌
(1/2) screen.
*Note (1) It is important to have an accuracy in thermal sensor when measuring center temperature of
the chamber.
(2) Input range:1.0
䉝
~
䇲
1.0
䉝
. It is unacceptable to input over the range at once.
<Display of Process 1>
<Display of Process 3>
<Display of Process 4>
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...