- 79 -
(2) H
2
O
2
generator (MCO-HP), H
2
O
2
Decontamination kit (MCO-HL)
z
Installation
Fill 80cc distilled water in MCO-HP. By using the cable enclosed, connect MCO-HP with the connector
on the bottom right of the incubator. (See picture as below)
Connector cap should be kept in safe place.
Ensure 2 pins of MCO-HP should be aligned to the holes on the rear duct. If not so, it may be caused
poor in spray performance.
For the check of spray performance, hold door switch by the tape.
z
Operation check
# Operation
Key
Display/Remark
1
-----
“Top
Screen”
2 Press
H
2
O
2
key for 3 seconds.
H2O2
“H2O2 Decontamination”
(System check starts)
3
H2O2
Decontamination
screen
‘H2O2 Decon’ blinks
-----
(If error message is shown, refer
to Instruction Manual P.57.)
4 Ensure if the outer door is locked by
electric key.
-----
5 Open outer door by using key to check
spray performance is correctly obtained.
-----
Turn key toward the direction
when you locked and take key off
before door shut. You cannot turn
the key with the door shut.
H
2
O
2
generator
UV lamp
Connector
Ultrasound oscillator
Front
Front
H
2
O
2
generator
Rear duct
UV lamp
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...