(2) Automatic set temperature alarm
Setting range: Set tempe/-1.0
Υ
~+/-5.0
Υ
(Default setting: +/-1.0
Υ
)
When a chamber temperature deviates from alarm setting range, the digits displaying current
chamber temperature will start blinking.
The following message will be displayed after an alarm delay time elapses.
“Warning:High Temp” --- For high temperature alarm
“Warning:Low Temp” --- For low temperature alarm
Buzzer will sound intermittently and remote alarm will be active.
(3) Automatic set CO2 density alarm
Setting range: Set CO2 d/-0.5%~+/-5.0% (Default setting: +/-1.0%)
When a CO2 density deviates from alarm setting range, the digits displaying current CO2 density
will start blinking.
The following message will be displayed after an alarm delay time (0~15minutes) elapses.
“Warning:High CO2 Density” --- For high CO2 density alarm
“Warning:Low CO2 Density” --- For low CO2 density alarm
Buzzer will sound intermittently and remote alarm will be active.
(4) Automatic set O2 density alarm
Setting range: Set O2 d/-0.5%~+/-5.0% (Default setting: +/-1.0%)
When a O2 density deviates from alarm setting range, the digits displaying current O2 density
will start blinking.
The following message will be displayed after an alarm delay time (0~30minutes) elapses.
“Waring:High O2 Density” --- For high O2 density alarm
“Warning:Low O2 Density” --- For low O2 density alarm
Buzzer will sound intermittently and remote alarm will be active.
(5)
Auto
Return
If there are no key operations for 90seconds in each setting screen, the setting mode will be
forcibly to the end and Top screen will be appeared.
(6) Key Lock
When Key Lock is active, the following message will be displayed on the stand-by setting screen.
“Key Lock” --- Key Lock is active
It is not permitted to change a stand-by setting value.
(7)
Door
alarm
Setting range of door alarm delay time: 1~30 minutes (Default setting: 2 minutes)
If an outer door is left opened, the following message will be displayed on the status screen.
“Door:Open” --- Outer door is left opened
Buzzer will sound intermittently after a door delay time elapses.
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...