3.
The Date Time Display will appear. Set the date, time and log interval.
・
Entering the date
Example for October 1, 2010: Enter 101001 in the Date Field.
・
Entering the time
Example for 10:05:00: Enter 100500 in the Time Field.
・
Entering the log interval
Example for 6 minutes: Enter 6 in Log Interval Field.
Note:
Ŷ
The default is 6 minutes.
Ŷ
The log interval can be set to between 2 and 30 minutes.
Ŷ
Relation between the log interval and the period that can be saved
1: Log interval of 2 minutes -- About 5 days
2: Log interval of 6 minutes -- About 14 days
3: Log interval of 30 minutes -- About 70 days
For the data beyond the above period, the older data is deleted to save the newer data.
1
D
a t e T
i m
e
3
D
a t e 1 0 / 1 0 / 0 1 ( Y Y / M M / D D )
4
T i m e
1 0 : 0 5 : 0 0
( h h
:
m m
:
s s )
5
L
o
g
I n
t e r v a
l
6 m i n
( 2 - 3 0 m i n )
6
- 146 -
39