32 SCANNER EMC INFORMATION
User’s manual
Planmeca Chairside CAD/CAM solution 165
32 SCANNER EMC INFORMATION
32.1 Guidance and manufacturer’s declaration – electromagnetic emission
The Planmeca PlanScan scanner is intended for use in the electromagnetic environment
specified below. The customer or the user of Planmeca PlanScan scanner should assure
that it is used in such an environment.
32.2 Guidance and manufacturer’s declaration – electromagnetic immunity
The Planmeca PlanScan scanner is intended for use in the electromagnetic environment
specified below. The customer or the user of Planmeca PlanScan scanner should assure
that it is used in such an environment
.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Planmeca PlanScan scanner uses RF
energy only for its internal function.
Therefore its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment
RF emissions
CISPR 11
Class A
The Planmeca PlanScan scanner is
suitable for use in all establishments, other
than domestic and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonics emission
IEC 61000-3-2
Not applicable
Voltage fluctuations /
flicker emissions IEC
61000-3-3
Not applicable
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
environment
-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV Contact
+/- 8 kV Air
+/- 6 kV Contact
+/- 8 kV Air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%
Electrical fast
transient / burst
IEC 61000-4-4
+/- 2kV for power
supply lines
+/- 1kV for input/
output lines
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV differential
mode
2 kV common
mode
Not applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.