fi-5530C/fi-5530C2
IMAGE SCANNER
MAINTENANCE MANUAL
APPR.
TITLE
DRAW.
No.
Feb.23, 2007
05
Section 4-9-5
4-9-5. Replacing the US Sensor (In Upper unit)
As the dust-proof area in the scanner is opened by the following procedure, do not replace the following parts in a dusty
room.
Refer to
Section 6-6
for the part number of the replacement part.
<Removing>
(1) Remove Chute unit (ADF paper chute), Stacker unit and Upper cover by referring the Sections from
4-7-1
,
4-7-2
and
4-7-5
.
(2) Remove the Optical unit by referring to
Section 4-9-3
.
(3) There is the US Sensor under the Optical unit. Disconnect a connector of US sensor, and then remove a screw that
secures a bracket to remove the US sensor.
<Mounting>
(1) Follow the removing procedure in reverse. When attaching the bracket, confirm that the tab of the bracket is inserted to
the slot of the frame. The rotational position does not affect the sensing properties, but attach the sensor securely by the
bracket.
(2) After mounting the US sensor, run the Sensor testing by the Maintenance mode #1 described in
Section 5-1-2
.
US Sensor
US Sensor
Bracket
Tab of bracket
shall be inserted to
the slot