3-3-1. Scanner does not turn ON (No display on the operator panel)
Table 3-3-1
Item
No.
Check items
How/where to check
1
Does the same symptom occur after
turning OFF and ON the scanner?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
Are the AC cable and AC adapter
correctly connected?
---
3
Replace the AC cable and AC adapter
and see if the error is resolved.
---
4
Replace Panel PCA and see if the error is
resolved.
Refer to
Section 4-9-1
.
5
Replace Control PCA and see if the error
is resolved.
Refer to
Section 4-11
.
3-3-2. Scanning does not start
Table 3-3-2
Item
No.
Check items
How/where to check
1
Does the same symptom appear when
turning the scanner ON again?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
Check the items listed in the right
column.
•
Are the AC cable and AC adapter correctly connected?
•
Is there documents loaded on Chute unit (ADF paper
chute)?
•
Is ADF cover completely closed?
•
Is interface cable correctly connected?
•
Is SCSI ID correctly set?
•
If any temporary error or alarm is indicated, follow the
corresponding troubleshooting.
3-3-3. Scanned image is distorted
Due to loose connection of connectors, cut wire in cables or defective parts, scanned image may have regular or random pattern
distortion on it.
Table 3-3-3
Item
No.
Check items
How/where to check
1
Check the items listed in the right
column.
•
Is interface cable correctly connected?
•
If any temporary error or alarm is indicated, follow the
corresponding troubleshooting.
2
Are the cables between Control PCA and
Optical unit damaged? Or are the
connectors connected correctly?
ADF front scanning: See
Section 4-10-6
for checking.
ADF back scanning: See
Section 4-9-3
for checking.
3
Replace Optical unit and see if the error
is resolved.
ADF front scanning: See
Section 4-10-6
for replacement.
ADF back scanning: See
Section 4-9-3
for replacement.
4
Replace Control PCA and see if the error
is resolved.
Refer to
Section 4-11
.