Section 3-3-16
3-3-16. (Reserved)
3-3-17. “E6: Operator panel alarm”
Table 3-3-17
Item
No.
Check items
How/where to check
1
Does the same symptom occur after
turning OFF and ON the scanner?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
It seems the installed Panel PCA has
been used in other scanner before and it
may be faulty one. Replace it with the
new Panel PCA.
If the EEPROM data is saved, replace the Panel PCA with new
one, and run Maintenance mode #7 by referring to
Section
5-1-8
.
3-3-18. “E7: EEPROM alarm”
Table 3-3-18
Item
No.
Check items
How/where to check
1
Does the same symptom occur after
turning OFF and ON the scanner?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
Replace the Panel PCA and see if the
error is resolved.
Refer to
Section 4-9-1
for replacement.
3
Replace the Control PCA and see if the
error is resolved
Refer to
Section 4-11
for replacement.
3-3-19. “E8: SCSI fuse blown”
Table 3-3-19
Item
No.
Check items
How/where to check
1
Does the same symptom occur after
turning OFF and ON the scanner?
Press power button of the scanner for more than 2 seconds to
turn it OFF, and after more than 2 seconds elapse, press the
power button to turn the scanner ON.
2
Check if the SCSI cable or other SCSI
devices connected to the scanner cause
this error.
The scanner can work even after this error occurs.
Go to item 3 if you want to remove this error.
3
Replace the Control PCA and see if the
error is resolved
Refer to
Section 4-11
for replacement.