Section 4-10-2
4-10-2. Replacing the HK Ring SR
As the dust-proof area in the scanner is opened by the following procedure, do not replace the following parts in a dusty room.
Refer to
Section 6-3
for the part number of the replacement part.
<Removing>
(1) Remove Chute unit (ADF paper chute), Stacker unit, PCB unit, Rear cover and Guide A ASSY by referring to Sections
from
4-7-1
to
4-7-4
and
4-7-6
.
(2) Slide the shaft of the plastic roller in the direction of
arrow (see right), and rotate the shaft to make flat
portion of the shaft vertical. Take out the roller and
shaft assembly.
(3) Remove two plastic rollers from the shaft, and remove
HK ring SR (4 pieces) from both ends of plastic rollers.
Handle HK ring carefully as it is easy to break.
<Mounting>
Follow the removing procedure in reverse. If the glasses in ADF are dirty, clean them by a dry lint-free cloth.
HK Ring SR
When attaching plastic rollers and
shaft, the flat portion of shaft shall
face following direction.
Be careful not to loose
this spring.
Flat portion of shaft
Plastic rollers