NewFEIMAS
Refer to Revision Record on page 2.
T.Anzai
Nov. 20, 2008
K.Okada
Rev
04
03
PAGE
39
/
138
PFU LIMITED
P1PA03334-B30X/6
Refer to Revision Record on page 2.
T.Anzai
July 7, 2008
K.Okada
Refer to Revision Record on page 2.
K.Okada
K.Okada
DESCRIPTION
CHECK
DATE
T.Anzai
APPR.
K.Okada
K.Okada
Aug.19, 2005
DESIG
N
DESIG
.
CUST.
I.Fujioka
I.Fujioka
T.Anzai
CHECK
fi-5530C/fi-5530C2
IMAGE SCANNER
MAINTENANCE MANUAL
APPR.
TITLE
DRAW.
No.
Feb.23, 2007
05
Section 3-1-1
Chapter 3 Troubleshooting
3-1 Self-diagnosis function
3-1-1. Scanner status display and self-diagnostics at power-on
(1) Operator panel display sequence at power-on
The following display is shown during initial processing.
Function No. Display
Power LED
Description
ON
Displays “8” without blinking.
Immediately after power-on, scanner turns all the segments ON.
When the initial processing starts, the following display appears.
Function No. Display
Power LED
Description
ON
Displays “P” without blinking.
It indicates scanner is currently in initial processing.
When the lamp intensity is getting close to the standard value, the following display appears.
Function No. Display
Power LED
Description
ON
Displays “0” without blinking.
It indicates the lamp intensity is getting close to the standard value.
When the initial processing terminates properly, the following display appears.
Function No. Display
Power LED
Description
ON
Displays default Function No. without blinking.
The scanner is in Ready state. The function No. “1” (default) is displayed.
The Function No. increments by 1 every time the Function button is pressed. After Function No. 9 is displayed, the number
changes to “C” and then returns to “1”.