5
Unpacking and Installing
This section applies to AURICAL Aud as well as to the various accessories available. If the accessories are not part of your
configuration, disregard the instructions relating to the accessories.
To install and get started with AURICAL Aud and the OTOsuite Audiometry Module, follow the sequence below:
•
Install OTOsuite on the PC before you connect to AURICAL Aud from the PC.
•
Unpack AURICAL Aud (see
).
•
Assemble AURICAL Aud (see
).
•
Connect the cables of the test equipment. See
–
–
AURICAL speaker unit installation
•
Run the OTOsuite Configuration Wizard to connect to and set up communication with AURICAL Aud. (See
.)
5.1
Requirements to the location
A sound cabin or sound treated room is not necessary, but it is recommended that you use a room where reverberation
time is not too long.
To ensure safe performance, AURICAL Aud must be correctly installed and the requirements listed in
and
must be complied with.
Warning
•
The charger unit should be kept away from the client area.
Warning
•
Keep AURICAL Aud away from all liquids and sources of heat (for detailed specifications, see
).
Warning
•
It is recommended to install the unit in an environment that minimizes the amount of static electricity.
For example, anti-static carpeting is recommended.
5.1.1
Requirements to noise in the test environment
Audiometry testing does not require a quieter testing environment than do other traditional Real Ear Measurement sys-
tems.
Although Audiometry testing is designed for use in a regular dispensing office environment, the test environment should
in some cases be adapted to suit the specific test types.
•
Make sure that the test environment is as quiet as possible. The quieter the room is, the more accurate your testing
will be.
It is very important that the measurements are not influenced by sounds or noise relating to anything else than the
presented test signals. Measurements of peak levels will be influenced by any sounds louder than the presented test
signals. Even extremely brief sounds may impact the measurements.
It is therefore important that the environment is well controlled to prevent any unforeseen influence.
•
Check that testing is not being done under an air conditioner or in front of a fan or ventilator.
Otometrics - AURICAL Aud
85
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...