Frequency
(Hz)
Nominal response level
(dB re. 1kHz level)
Tolerance
(dB)
3000
-15.5
±4
4000
-11.0
±6
Operator accessories
Operator monitor headset - headphones: •
40 mW 16
Ω
•
3.5 mm (1/8 inch) stereo jack
Operator microphone (desktop or boom): •
Electret microphone
•
Input voltage: 0.002 to 0.02 Vrms,
•
Input resistance: 2.21 k
Ω
.
•
3.5 mm (1/8 inch) jack
USB port connector
Type:
USB device port
Compliant:
USB 2.0
Speed:
High speed
Transport and storage
Temperature:
-30°C to +60°C (-22°F to 140°F)
Air humidity:
10% to 90%, non-condensing
Air pressure:
500 hPa to 1060 hPa
Operating environment
Mode of operation:
Continuous
Temperature:
+15°C to +35°C (59°F to 95°F)
Air humidity:
30% to 90%, non-condensing
Air pressure:
700 hPa to 1060 hPa.
(Operation in temperatures exceeding -20°C (-4°F) or +60°C (140°F) may cause permanent damage.)
Warm-up time
< 5 min.
Note
•
Should be extended if AURICAL Aud has been stored in a cold environment.
126
Otometrics - AURICAL Aud
13 Technical specifications