13.5
Notes on EMC (Electromagnetic Compatibility)
•
AURICAL Aud is part of a medical electrical system and is thus subject to special safety precautions. For this reason, the
installation and operating instructions provided in this document should be followed closely.
•
Portable and mobile high-frequency communication devices, such as mobile phones, may interfere with the func-
tioning of AURICAL Aud.
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
AURICAL Aud is intended for use in the electromagnetic environment specified below. The user of AURICAL Aud should ensure that it is used in such an envir-
onment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
AURICAL Aud uses RF energy only for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
AURICAL Aud is suitable for use in all environments, including domestic environments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
AURICAL Aud is intended for use in the electromagnetic environment specified below. The user of AURICAL Aud should ensure that it is used in such an envir-
onment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 6 kV con-
tact
+/- 8 kV air
+/- 6 kV con-
tact
+/- 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Otometrics - AURICAL Aud
129
13 Technical specifications
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...