![otometrics AURICAL Aud Reference Manual Download Page 65](http://html.mh-extra.com/html/otometrics/aurical-aud/aurical-aud_reference-manual_1658652065.webp)
2. Slide the probe cable into place so it fits snugly
in the cable holder on an Ambient Noise
Assessor clip. Repeat with another clip for the
second probe.
3. Place FreeFit on the patient's shoulders. Clip
the probes to the FreeFit collar.
–
Position the probes so that they are above
the patient's shoulders.
4. Adjust each probe carefully:
–
The reference microphone should face the
ceiling.
A. Reference
microphone
–
Ensure that the probe housing does not
touch the patient's clothing and that the
microphone is not covered by anything,
such as clothing or hair.
5. Open the audiometry control panel.
If the
Ambient Noise Assessor
is not already open, click the button on the Audiometry toolbar.
6. Press the power button on FreeFit. If FreeFit does not connect, ensure that your FreeFit has been selected in the
PMM configuration wizard. Also check the battery status of FreeFit. (See Charging, Powering and Connection Status in
the reference manual for FreeFit.)
The
Ambient Noise Assessor
dialog indicates the status of the noise level:
Ambient Noise Status
The measured ambient noise level is below the maximum permissible level for the stim-
ulus that is currently selected.
Otometrics - AURICAL Aud
65
2 Navigating in the OTOsuite Audiometry Module
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...