How to score
•
Use the score buttons or shortcut keys (
+
,
-
), or make a right click on a given word already played in the
Word list con-
trol
.
When to reset score
Notice the field
Reset Score Counter
in the
Options
dialog.
•
Reset Score Counter
=
True
The
Scoring panel
as well as the
Word list control
images (
+
,
-
) is reset if changes are made to the
Channel Settings
(except
Continuous On
), the
Intensity
or when the
Reset
button is clicked.
•
Reset Score Counter
=
False
We actually reset the score in some cases even if
Reset Score Counter
=
False
. This is in situations where we go from
using a word list to e.g.
Stimulus CH1
=
Stimulus CH2
=
Mic.
4.4
Performing sound-field speech audiometry off site
When a service technician performs an annual calibration on your AURICAL Aud speaker, the calibration is performed in
the location where you most often use AURICAL Aud, for example, in your consultation office. However, you may want to
take AURICAL Aud with you to perform sound-field speech audiometry in other locations, such as in a client's home or in a
nursing home. In that case, the AURICAL Aud speaker must be calibrated for the new location. You can use a sound level
meter, together with OTOsuite, to make a temporary manual calibration.
Overview of portable sound-field audiometry tasks
To prepare for off-site sound-field speech audiometry:
•
Enable portable sound-field audiometry in the Configuration Wizard.
•
When you are at the site where you will perform off-site sound-field audiometry, click
Calibrate
in the Speech Audi-
ometry control panel to open the
Portable Sound-Field Audiometry
window. Use this window to:
–
Calibrate one speaker - first the primary signal, then the secondary signal
–
Calibrate the second speaker, if applicable. First the primary signal, then the secondary signal
–
Save the calibration values.
•
To use the new calibration values, select
Portable Sound-Field Audiometry
in the Control panel, and then perform
speech audiometry as you normally would.
If you move to another off-site location, remember to repeat the manual calibration.
When you return to your default location, deselect
Portable Sound-Field Audiometry
in the control panel.
4.4.1
Enabling portable sound-field audiometry
Before you can make a manual calibration of the speaker, you must:
•
Select
Enable Portable Sound-Field Audiometry
in the configuration wizard, on the
Preferences
page.
Once you have enabled portable audiometry, the
Portable Sound-Field Audiometry
option appears in the control panel
on the Speech tab.
4.4.2
Performing a manual calibration of AURICAL Aud speakers
To prepare for portable sound-field audiometry, you must configure the speakers and then calibrate them.
Otometrics - AURICAL Aud
83
4 Examples of audiometric testing
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...