8. If air conduction perception is better than bone conduction, this is a positive Rinne and agrees with normal hearing
9. If bone conduction perception is better than air conduction, this is a negative Rinne and indicative of a conductive
hearing loss.
10. If air conduction and bone conduction are equal, this is a positive Rinne and indicative of a sensorineural hearing loss.
11. Store the test result in the
Rinne
Box.
4.2.6
Performing Alternate Binaural Loudness Balancing (ABLB) test
If you have measured thresholds prior to performing ABLB, then these will appear already plotted in the ABLB graph.
1. Properly place the earphones.
2. Give the patient the responder(s).
3. Instruct the patient so that he/she knows what to expect: "You will hear a continuous tone. The level of this tone will
be changed periodically. Press the button on the responder immediately each time you hear a change. I will ask you if
the test tone is softer than, louder than or equal to the reference tone in the opposite ear."
4. Choose the ABLB tab or select ABLB in the Test Selector.
–
Make sure that the control panel of the OTOsuite software is set to the correct transducer and desired routing.
–
If left ear is selected as the test ear in the Control Panel Routing section, then the right ear will be the reference
ear.
5. Set the reference level (
Ch 1
).
6. Set the test level (
Ch 2
) to 10 dB above the patient's threshold for that ear.
7. Adjust the level of the test signal (
Ch 2
) until the patient reports that the two signals sound equally loud.
8. Store the data points.
9. Repeat steps
to
for additional frequencies.
10. If desired, switch routing to test the opposite ear and repeat steps
to
.
4.2.7
Performing Short Increment Sensitivity Index (SISI) test
1. Properly place the earphones.
2. Give the patient the responder(s).
3. Instruct the patient so that he/she knows what to expect: "You will hear a continuous tone. The level of this tone will
be changed periodically. Press the button on the responder immediately each time you hear a change."
4. Choose the
SISI
tab or select
SISI
in the Test Selector.
–
Make sure that the control panel of the OTOsuite software is set to the correct transducer and desired routing.
–
Make sure that the frequency is correct.
5. Begin the test in the ear with the lower PTA or SRT.
6. Start with 5 dB SISI dB step and set the level 20 dB above the patient's threshold for that frequency in that ear.
7. The test begins with 1000 Hz.
8. Make sure that the patient understands the response task.
9. Change the SISI dB step to 1 dB. Set the level 20 dB above the patient's threshold for that frequency in that ear and
press the
Play
button.
10. Store the data points.
11. Repeat steps
to
for additional frequencies.
Otometrics - AURICAL Aud
81
4 Examples of audiometric testing
Summary of Contents for AURICAL Aud
Page 1: ...AURICAL Aud Reference Manual Doc No 7 50 1030 EN 15 Part No 7 50 10300 EN ...
Page 12: ...1 Introduction to AURICAL Aud 12 Otometrics AURICAL Aud ...
Page 70: ...2 Navigating in the OTOsuite Audiometry Module 70 Otometrics AURICAL Aud ...
Page 100: ...5 Unpacking and Installing 100 Otometrics AURICAL Aud ...
Page 104: ...7 Communicating with the device 104 Otometrics AURICAL Aud ...
Page 114: ...11 Troubleshooting 114 Otometrics AURICAL Aud ...