Setting the Output Conditions
164
ZW-8000/7000/5000
User's Manual
Multi View Explore
: [Bank Group] | [(Bank Data Name)] (double click)
→
Edit pane
: [Task Settings] icon (
)
→
Task Settings window : [Hold]
Set “Trigger hysteresis” according to the fluctuation of the measured value near the trigger level. Hysteresis is applied
simultaneously with start of measurement to prevent chattering of the TIMING input.
Trigger
Set-
tings
Trigger method
Setting Valid value
The period where the measurement value is a valid value is taken to be the
sampling period
When the delay time is set, the timing when the measurement
value becomes invalid value and when the sampling period is
finished are not synchronized.
Timing when the Setting Valid value finishes is after the sampling
period passed.
Trigger level
-999.999999 to 999.999999
[mm]
(default value: 0)
Set the self-trigger level when [SELF-UP] or [SELF-DOWN] are selected at
[Trigger method].
Trigger
hysteresis
0 to 999.999999 [mm]
(default value: 0.05% of
measuring range)
Set the hysteresis width for the self-trigger when [SELF-UP] or [SELF-
DOWN] are selected at [Trigger method].
1
Set the hold mode.
2
Select the trigger method from [Trigger
method].
3
Enter [Trigger level] and [Trigger hysteresis]
when [Selfup trigger] or [Selfdown trigger]
are selected at [Trigger method].
The trigger can also be set by the operating keys on the Sensor Controller.
Item
Setting item
Setting value
Description
Me
asu
rement v
a
l
u
e
i
s
cl
a
mp
s
t
a
t
us
Me
asu
rement
v
a
l
u
e
Sa
mpling period
Sa
mpling period
Me
asu
rement v
a
l
u
e
i
s
cl
a
mp
s
t
a
t
us
Me
asu
rement v
a
l
u
e
i
s
cl
a
mp
s
t
a
t
us
Oper
a
te point
Ret
u
rn point
Important
Important
Note
Summary of Contents for ZW-7000 Series
Page 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Page 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Page 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Page 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Page 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Page 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Page 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Page 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Page 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Page 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Page 335: ......