Setting the Output Conditions
ZW-8000/7000/5000
User's Manual
161
4
Settin
gs for Function
Multi View Explore
: [Bank Group] | [(Bank Data Name)] (double click)
→
Edit pane
: [Task Settings] icon (
)
→
Task Settings window : [Hold]
Hold
Hold mode
Auto Peak to
Peak
Holds the difference between the maximum and minimum values of the
measurement result. Output the measurement value at each updating of the
difference value Maximum and Minimum.
Average
Holds the “average of the measured value” of the sampling period. The current
measurement value is output after Sampling, and then this state is held until
the next sampling ends.
Sampling
Holds the measured value the moment that control enters the sampling
period. The current measurement value is output after Sampling, and then this
state is held until the next sampling is started.
Through
(default value)
The hold function is not used. The measured value is always output.
1
Select the hold mode from [Hold mode].
Setting items for required parameters are
displayed depending on the selected hold
mode.
The hold mode can also be set by the operating keys on the Sensor Controller.
Item
Setting item
Setting value
Description
0
0
Current
measured value
Output
(Max.value - Min.value)
Current
measured value
Output
(Average)
Sampling period
Current
measured value
Output
Sampling period
Note
Summary of Contents for ZW-7000 Series
Page 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Page 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Page 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Page 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Page 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Page 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Page 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Page 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Page 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Page 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Page 335: ......