Performing internal logging
116
ZW-8000/7000/5000
User's Manual
3
After internal logging ends, click the [Save to file (Sensor
→
PC)] to output the data to a file.
When saving to a file, set the saving start and end points on the following screens, and then press the
"Save" button.
A CSV format file in the following format is output. Unit of each output values are the following:
Measure value: nm
Amount of emitted light: 0.01% (0.01 to 100.00%)
Amount of received light: Luminance level (0 to 4095)
LoggingMode
Internal
Index
OUT1
OUT2
OUT3
OUT4
0
-1938
2823
9437
-2147483648
1
-1939
2764
9401
-2147483648
2
-1940
2773
9355
-2147483648
3
-1940
2807
9523
-2147483648
4
-1941
2794
9442
-2147483648
5
-1941
2817
9497
-2147483648
• L
ab
el in
s
ert mode: OFF
• L
ab
el in
s
ert mode: ON
Summary of Contents for ZW-7000 Series
Page 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Page 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Page 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Page 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Page 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Page 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Page 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Page 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Page 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Page 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Page 335: ......