Part Names and Functions
62
ZW-8000/7000/5000
User's Manual
ECAT status indicator LED
LED name
Color
Status
Contents
ECAT RUN indicator
Green
OFF
Initialization status
Blinking
Pre-Operational status
Single flash
Safe-Operational status
ON
Operational status
ECAT ERROR indicator
Red
OFF
No error
Blinking
Communication setting error or PDO mapping error
Single flash
Synchronization error or communications data error
Double flash
Application WDT timeout
ON
PDI WDT timeout
L/A IN indicator
Green
OFF
Link not established in physical layer
Flickering
In operation after establishing link
ON
Link established in physical layer
L/A OUT indicator
Green
OFF
Link not established in physical layer
Flickering
In operation after establishing link
ON
Link established in physical layer
The timing of the flashing of the EtherCAT operation indicators is as follows:
Note
ON
OFF
ON
OFF
ON
OFF
ON
OFF
Flickering
Blinking
Single flash
Double flash
50 ms
200 ms 200 ms
200 ms
200 ms
1000 ms
200 ms 200 ms 200 ms
1000 ms
Summary of Contents for ZW-7000 Series
Page 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Page 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Page 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Page 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Page 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Page 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Page 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Page 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Page 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Page 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Page 335: ......