Setting the Temperature with the Control Panel
To enable the incubator, press the
Temp on/off
button on the control
panel. The display indicates that temperature control is on by displaying
numbers in the lower half of the LCD. The instrument sets the reading
chamber to the default temperature, 37°C.
Figure 3-6:
Temperature Control Enabled
To change the temperature set point, press the up or down arrow keys
until the desired temperature set point appears on the display.
The reading chamber temperature is maintained at the set point until
you disable temperature control by touching the
Temp on/off
button
again. When the incubator is off, the temperature within the reading
chamber returns to ambient.
Note:
If the power is shut off to the instrument for any reason, you
need to turn on the incubator again and allow sufficient time (at least
10 minutes) for the control algorithm to fully stabilize the reading
chamber temperature.
FlexStation_manual.book Page 73 Tuesday, June 25, 2013 10:22 AM