FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
54
0112-0127 B
To install the fluidics module:
1.
Grasp the handle on the fluidics module and use a second person
to help to lift the module into position over the detection
module.
2.
Tilt module up at about 90° and have one person hold it in
position while the other person aligns the quarter-turn fasteners
in the flange to the holes in the bottom of the fluidics module.
3.
Connect the quarter-turn fasteners to the bottom of the fluidics
module and lock the fasteners into place.
4.
Attach the restraining cable from fluidics module to the right-
most mounting tab on the top cover of FlexStation 3 instrument.
The restraining cable comes attached to the fluidics module.
Figure 2-3:
Restraining cable attached to the fluidics module
Attach the other end of the restraining cable the right-most
mounting tab on the top cover.
Figure 2-4:
Restraining cable attached to the top cover
FlexStation_manual.book Page 54 Tuesday, June 25, 2013 10:22 AM