Troubleshooting Procedures
0112-0127 B
151
Removing the Fluidics Module
1.
Turn the power off to the instrument. Disconnect the power
cords from the instrument and the cables connected to the
computer.
2.
Press in the handle at the bottom of the top cover and lift the
cover back off the detection module, revealing the fluidics
module.
3.
Using the handle, tilt the fluidics module to the left.
Figure 5-9:
Tilting the Fluidics Module
4.
Locate the electrical connector between the bottom of the
fluidics module and the housing, and then unscrew the retaining
screws and unplug the connector.
5.
Locate the fluidics module restraint (a green cable) and
disconnect the restraint from the housing.
CAUTION!
To avoid damage to the instrument, follow these instructions
and any instruction labels on the instrument exactly.
CAUTION!
Never operate the instrument with the covers
removed. Damage to the instrument can occur. Read and obey
all warning labels.
FlexStation_manual.book Page 151 Tuesday, June 25, 2013 10:22 AM