0112-0127 B
155
B
Performance Specifications
Thermal specifications for microplates used in the FlexStation 3
instrument apply to flat-bottom microplates with isolated wells. All other
microplate specifications apply to standard 96-well polystyrene
flat-bottom microplates.
When pathlength compensation is applied to microplate absorbance
measurements, the solution volume in the microplate well is between
100
μ
L and 300
μ
L.
Technical specifications are subject to change without notice.
Table B-1:
Fluidics Specifications for the 8-Channel Pipettor
Item
Specification
Volume range
1
μ
L to 200
μ
L
Dispensing precision
8% CV at 5
μ
L
2% CV at 50
μ
L
Maximum dispense rate
208
μ
L per second
Table B-2:
Fluidics Specifications for the 16-Channel Pipettor
Item
Specification
Volume range
1
μ
L to 30
μ
L
Dispensing precision
3% CV at 10
μ
L
5% CV at 1
μ
L
Maximum dispense rate
52
μ
L per second
FlexStation_manual.book Page 155 Tuesday, June 25, 2013 10:22 AM