Troubleshooting Procedures
0112-0127 B
149
Tilting or Removing the Fluidics Module
If you need to tilt or remove the fluidics module to clean up after a large
spill, or to recover tips, follow these instructions.
Tilting the Fluidics Module
1.
Turn the power off to the instrument. Disconnect the power
cords from the instrument and the cables connected to the
computer.
WARNING! The following procedures must be completed by a
service-trained user. Do not attempt the following procedures if you
have not been trained properly by appropriate Molecular Devices
personnel.
WARNING! Turn off the power switch and disconnect the power cord
from the main power source before performing any maintenance
procedure that requires removal of any panel, cover, or disassembly of
any interior instrument component.
WARNING! BIOHAZARD. This procedure includes steps involving
potentially infectious biological agents requiring that proper handling
precautions be taken.
CAUTION!
Never operate the instrument with the covers
removed. Damage to the instrument can occur. Read and obey
all warning labels.