FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
116
0112-0127 B
Cleaning Up Spills
If fluids spill into the bottom of the reading chamber when the reading
chamber drawer is outside the instrument, the fluid is directed to a tray
at the bottom of the instrument, from which it exits to the bench or
counter beneath the instrument. Wipe up spills immediately. Clean the
exterior of the unit and the drawer as necessary.
Cleaning the Fan Filter
The fan filter on the bottom of the reader requires periodic cleaning. The
frequency of the cleaning depends on how dusty your particular lab is
and could range from once per month to once every six months.
1.
Remove any microplates, adapters, reservoirs, and tips from all
three drawers, and then close the drawers.
2.
Turn the power to the reader
OFF
and then remove the power
cord and cables from the back of the reader.
3.
Turn the reader over so that it rests flat on the bench.
4.
Remove the black fan cover and then remove the filter.
5.
Clean the filter by blowing canned air through it or by rinsing it
first with water and then with alcohol, and allowing it to dry
completely.
6.
Place the clean, dry filter over the fan and replace the black
cover.
7.
Turn the reader back over.
8.
Reconnect the power cord and cables to the reader and return
the plate adapters to their respective positions.
WARNING! Always make sure the power switch on the instrument is
in the OFF position and remove the power cord from the back of the
instrument before opening any covers or panels.
CAUTION!
Do not attempt to assemble or disassemble the
instrument with the pipette tips or compound and read plates in
place. Spillage or damage to the pipette tips, plates, or the
instrument can occur.
FlexStation_manual.book Page 116 Tuesday, June 25, 2013 10:22 AM