Installation
0112-0127 B
51
Unpacking the Fluidics Module and Accessories
1.
Remove the fluidics module from the cardboard box inside the
wooden crate and take it out of the protective bag. Set it in a safe
place.
2.
Remove the box (containing the pipettor head) and the bags of
accessories.
3.
Open the accessories bags and remove cables and the hex key.
You will need them later in the assembly procedure.
4.
Set packaging aside.
Unpacking the Detection Module
Tools Needed
•
Hex key, 3/32 inch ball drive, L (provided)
•
Phillips screwdriver (not provided)
To unpack the detection module:
1.
Use two people to unlatch the midsection of the crate (on top
and bottom) and then move that midsection aside. Due to the
size of the crate, this step requires two people.
2.
Slide the plastic bag enclosing the instrument out of the way,
around the base of the detection module.
3.
Use two people to reach inside the bag and under the
instrument and then lift the instrument out of its shipping tray
and place it on the bench.
4.
Set the shipping materials out of the way.
Note:
Keep the system in a location that is dedicated to its use, on a
level surface, away from direct sunlight, dust, drafts, vibration, and
moisture.
WARNING! The instrument weighs approximately 50 pounds
and should be lifted with care. To prevent injury, use at least
two people to lift the instrument.
FlexStation_manual.book Page 51 Tuesday, June 25, 2013 10:22 AM