Troubleshooting Procedures
0112-0127 B
131
There are two settings in the Preferences dialog that must be
correct for the instrument to operate.
4.
Make sure the Serial Port selected is the same as the one on the
computer to which the instrument is physically connected. It
might be COM1 or COM2.
5.
Make sure the Serial Comm Speed is correctly selected. The
communication speed is 9600 during the instrument’s initial
warm up sequence, while it is first communicating with the
software.
If you continue to see the red X after confirming the Serial Comm Speed,
see
Recovery Procedure on page 146
.
If you follow these steps and are still having trouble starting up the
system, contact Technical Support. See
Note:
The Serial Comm Speed automatically changes to 57,600
when you select Flex mode from the Settings window.