FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
144
0112-0127 B
7.
Pull the spiral cord off the coil hook behind the pipettor head.
This will free the mechanism for easier removal.
Figure 5-6:
Pipettor Head Coil Hook
Note:
Note: If you want to dispense liquid from the tips, skip the
next two steps and see next
Expelling Undispensed Fluid from
8.
To completely remove the pipettor head, unscrew the black
connector at the end of the white spiral cord from its secure
position around the receptacle.
9.
Pull the connector out of the receptacle and take the pipettor
head the rest of the way out of the fluidics module.
FlexStation_manual.book Page 144 Tuesday, June 25, 2013 10:22 AM